Direct comparison of the FibroScan XL and M probes for assessment of liver fibrosis in obese and nonobese patients

2013 Hepatic Medicine: Evidence and Research 5;(43-52)

BACKGROUND: A novel Fibroscan XL probe has recently been introduced and validated for obese patients, and has a diagnostic accuracy comparable with that of the standard M probe. The aim of this study was to analyze and understand the differences between these two probes in nonobese patients, to identify underlying causes for these differences, and to develop a practical algorithm to translate results for the XL probe to those for the M probe. METHODS AND RESULTS: Both probes were directly compared first in copolymer phantoms of varying stiffness (4.8, 11, and 40 kPa) and then in 371 obese and nonobese patients (body mass index, range 17.2-72.4) from German (n = 129) and Canadian (n = 242) centers. Liver stiffness values for both probes correlated better in phantoms than in patients (r = 0.98 versus 0.82, P 10 kPa for F0, F1-2, F3, and F4 fibrosis, respectively) significantly improved agreement between the two probes from r = 0.655 to 0.679. CONCLUSION: Liver stiffness can be measured in significantly more obese and nonobese patients using the XL probe than the M probe. However, the XL probe is less accurate and adjusted cutoff values are required.