Assessment of renal allograft fibrosis by transient elastography

2013 Transplant International 26;5 (545-551)

Transient elastography (TE, Fibroscan) has been established as a noninvasive assessment tool of liver fibrosis. We evaluated potentials and limitations of TE for identifying renal allograft fibrosis. The technical possibility of kidney examination by TE was assessed in two 10-week-old German landrace pigs and kidney stiffness (KS) was evaluated in 164 renal transplant patients. KS could be determined in all animals at the pole and pars media (29 +/- 10 kPa vs. 31 +/- 17 kPa). In human renal allografts KS was successfully performed in 94.5% of the test series with reliable results in 72% of the measurements. Mean KS at the pole or pars media were comparable (35.0 +/- 19.9 kPa vs. 33.2 +/- 18.6 kPa). Significantly higher KS was detected in renal allografts with histologically confirmed advanced fibrosis. Body-mass-index, skin-allograft distance, and peri or intrarenal fluid accumulation were important confounders of successful KS measurements (BMI: r = -0.31; P